New Video – How to Replace the Vac-chip™ on Isolera™ Dalton

30 September 2015

The vac-chip™ on Isolera™ Dalton is a vacuum interface. It allows the mass detector to achieve the operating pressures required for analysis and it aligns the ions from the spraychip® into the vacuum system through the vac-chip. The vac-chip may become blocked or dirty over time and it may need to be cleaned or replaced. This 3D-animation will show you step-by-step how to replace it.

 

Required: Powder-free gloves and a vac-chip™ (P/N 413712SP).
You also need the pair of tweezers, flat headed screwdriver, Allen key, and dust cover supplied with the mass detector.
1. Press Main Menu in the right-hand panel.
2. Press System. The Select User dialog opens.
3. Select your user name and press OK.
4. If your account is password-protected, enter your password and press OK.
5. Select the Maintenance tab.
6. Press replace vac-chip in the Mass Detector field and follow the instructions on the screen.