The RapidTrace®+ was withdrawn on July 22nd 2019
Please contact us here to discuss alternative products or why not check out Biotage® Extrahera™
Waste outlets | 4 separate |
Gas inlet (optional) | 1-gas inlet, 2.8 bar maximum |
Connections | RS232 Input: Connects the module to the controller RS232 Output |
Dimensions | 10 cm x 55 cm x 60 cm (W x D x H) |
Weight | 14,5 kg |
Power Requirements | 50–60 Hz, 100–240 VAC ±10% |
Column sizes | 1, 3 or 6 mL |
Sample size | 0.1 mL to 14 mL C50000 – RapidTrace+ Workstation 3 mL (0.1 to 40 |
Syringe pump | Load Volume 0.1 mL to 5.8 mL |
Flow rates | 0.1 mL/min to 42 mL/min (0.006 mL/sec to 0.7 mL/sec) |
Liquid Handling Accuracy | Volumes 3.0 mL ± 1% |
Mixing Capability Cycle | Maximum mixing volume of 5.0 mL in mixer |
Liquid Sensing | Volumes of 1.2 mL or larger |
Solvent inlets | 8 reagent inlets, individually configurable aspirate/dispense ra |
Cyclosporine is an immunosuppressant used in organ transplants to overcome rejection. Following the transplant of a heart, liver, kidney or lung, a patient must be regularly tested for the rest of their lives. Blood testing intervals vary from 1 month to 6 months and this rapid reference method is ideal for batch tests with the advantage of no metabolite interfer-ence which is always found in immunological methods.
Tags: Application Notes,
Liver
The removal of interferences and sample concentration are
essential steps in trace level detection of pesticides, sheep
dips, herbicides and PCBs in river water, wastewaters and
trade effluents. The following methods were developed by
the Environment Agency National Laboratory Service facility
at Nottingham using the Biotage RapidTrace for automated
solid phase extraction and the TurboVap II for evaporation.
The procedures are UKAS accredited and can reach detection
limits of 0.001 ug/Litre for some compounds.
Tags: Application Notes,
Environmental,
Environmental contaminants,
Organophosphates,
RapidTrace,
Water
With the widespread clinical use of antidepressant drugs comes the need to accurately measure therapeutic lev-els to determine proper dosage, as well as check for toxic levels and patient compliance. More and more laboratories are shifting their therapeutic drug monitor-ing from liquid extraction to solid phase extraction. A large clinical laboratory has automated the process of extracting these drugs from serum using the RapidTrace SPE Workstation.
Tags: Application Notes
EPA methods 8260 & 8270 cover the analysis of polychlorinated
dibenzo-p-dioxins and polychlorinated dibenzofurans
in ground & surface waters. This also now includes
1,4-dioxane and PCBs. This application note will focus on
the automated SPE of the sample rather than the detection
which usually involves GCMS or LCMS. NB The columns
and reagents used for soils and sediment analysis differ to
ground water extraction.
Tags: Application Notes,
Environmental,
Environmental contaminants,
RapidTrace,
Soil
The use of schedule I drugs for patient pain management therapy warrants constant monitoring of therapeutic levels in the patient. Screening patient urine samples for the free drugs is complicated by the metabolism process which converts the free drug to the -glucuronide form. Patient urine samples can be enzymatically hydrolyzed to cleave glucuronide moiety and produce the free form of drug. The target analytes can them be extracted from the urine matrix using EVOLUTE CX solid phase extraction cartridges. LOQs range from 5 ng/mL to 0.5 ng/mL with recoveries above 80% and RSD <10%
Pain Management, EVOLUTE, CX, SPE, Solid Phase Extraction, Polymer, LC-MS-MS, Cocaine, Diazepam, Heroin, Morphine, Oxycodone, Fentanyl, Diazepam, Mixed Mode, RapidTrace
Tags: Alprazolam,
Application Notes,
Buprenorphine,
Butylone,
Clinical,
Clonazepam,
Codeine,
Column,
Fentanyl,
Flunitrazepam,
Hydrocodone,
Hydromorphone,
LC-MS/MS,
Meperidine,
Methadone,
Morphine,
Naloxone,
Naltrexone,
Oxycodone,
Oxymorphone,
RapidTrace,
Temazepam,
Urine
Melamine is traditionally used in making plastics, however, it’s low cost and high nitrogen content has led to exploitation in various sections of the food industry, most notably involving dairy products. Sustained melamine exposure can result in kidney stones and renal failure, with the young being most susceptible. This application note has been optimized on the RapidTrace+; an updated version of the current RapidTrace software which allows the system to process smaller bed heights than were previously possible. The application note shows recoveries of 88-90% with RSDs <10%.
Melamine, food scare, china, EVOLUTE, cation exchange, RapidTrace, Automation, SPE, sample prep,
Tags: Application Notes,
Column,
Food,
Food contaminant,
LC-MS/MS,
Melamine,
Milk,
RapidTrace
Ethyl glucuronide (EtG, Figure 1.) is an ethanol metabolite formed by the glucuronidation of the parent molecule. The presence of EtG in urine and plasma is a useful and sensitive biomarker of alcohol intake in forensic toxicology. EtG can be detected at very low levels and is useful in distinguishing between ingested ethanol and that which is adsorbed through the skin from ethanol. This application note has been optimized on the RapidTrace+; an updated version of the current RapidTrace software which allows the system to process smaller bed heights than were previously possible. The application note shows analyte recoveries all above 97% with RSDs >10%.
EVOLUTE, Resin, EtG, Alcohol, Alcohol Biomarker, metabolite, anion exchange, Strata X, Oasis, Waters,
Tags: Alcohol metabolite,
Application Notes,
Column,
Ethyl glucuronide (EtG),
LC-MS/MS,
RapidTrace,
Urine
This application note describes an automated extraction procedure for the opiates, morphine and codeine from urine using ISOLUTE HCX and the RapidTrace Workstation.
Opiates, drug testing, DOA, Ion exchange
Tags: Application Notes,
Codeine,
Column,
Forensic,
Morphine,
Opiates,
RapidTrace,
Urine
Cyclosporine (Figure 1) is an immunosuppressant drug most significantly used in post-operative organ transplants to reduce the chance of rejection of donor organs by decreasing the activity of the patient’s immune system. Subsequent to an organ transplant, patients are tested at regular intervals for the presence of very low levels of the compound. Cyclosporine is a cyclic non-ribosomal peptide of 11 amino acids and contains a single D-amino acid.
Cyclosporine, Cyclosporine, Immunosuppressant,
Tags: Application Notes,
Clinical,
Column,
Cyclosporin,
Cyclosporine,
Immunosuppressant,
RapidTrace,
Whole blood
The typical range of drugs screened using this generic method are Cocaine, D3-Cocaine, Benzoylegonine-TMS, D3-Benzoylegonine-TMS, Amphetamine, D5-Amphetamine, Methamphetamine, Codeine, Morphine, Amitriptyline HCL, Nortriptyline HCL, Imipramine HCL, Desipramine HCL. These drugs represent a panel of the most significantly
abused drugs as recommended by The National Institute on Drug Abuse (NIDA).
drugs of abuse, drugs testing, amphetamines, BZE, Cocaine, Codeine, Ion exchange,
Tags: Amitriptyline,
Amphetamine,
Application Notes,
Benzoylecgonine,
Biological fluids,
Butylone,
Codeine,
Column,
Desipramine,
Forensic,
Imipramine,
Methamphetamine,
Morphine,
Nortryptiline,
Plasma,
RapidTrace,
Tricyclic Antidepressant,
Urine
This application note describes an automated extraction procedure from Tricyclic Antidepressants from Whole Blood Using ISOLUTE HCX Columns and the RapidTrace Workstation.
TCAD, TCA, Tricyclic antidepressants, ISOLUTE,
Tags: Amitriptyline,
Application Notes,
Clinical,
Column,
Desipramine,
Imipramine,
Nortryptiline,
RapidTrace,
Tricyclic Antidepressant,
Whole blood
This application note describes an automated extraction procedure for Cocaine metabolite, Benzoylecgonine from urine using ISOLUTE HCX and RapidTrace Workstation.
Tags: Application Notes,
Benzoylecgonine,
Cocaines,
Column,
Forensic,
RapidTrace,
Urine
This Application Note describes the operating conditions for the automated fractionation of EPH into aliphatic and PAH fractions using the ISOLUTE EPH column in conjunction with the RapidTrace SPE automation system.
EPH, Massachusetts Department of Environmental Protection,
Tags: Aliphatic hydrocarbos,
Application Notes,
Column,
Environmental,
Environmental contaminants,
Polyaromatic hydrocarbons,
RapidTrace,
Soil
When analyzing human urine for drugs of abuse, one of the most common tests is for the cocaine metabolite Benzoylecgonine (BZE). A contract laboratory has automated this labor intensive procedure using the RapidTrace.
Tags: Application Notes,
Benzoylecgonine,
Column,
Drugs of Abuse,
Forensic,
GC-MS,
RapidTrace,
Urine
The extraction and segregation of Organic compounds from core samples taken from marine sediments is used to predict climate conditions, reconstruct sea temperatures and identify the type of vegetation from the C3 ( Temperate cli-mate) and C4 ( Tropical Climate) isotopic mass spectroscopy ratio analysis.
Tags: Application Notes
This method details extraction of the NIDA-5 (SAMHSA-) drug of abuse classes using the same SPE column (ISOLUTE HCX 130 mg/1 mL) and single set of reagents (8 in total) on the RapidTrace.
Tags: Application Notes,
RapidTrace
The following method addresses the extraction of 6-monoacetyl morphine from urine using
non-polar and strong cation exchange retention mechanisms. This compound can confirm
heroine use since it is a heroine metabolite, but not a metabolite of codeine or morphine
Tags: Application Notes,
Codeine,
RapidTrace,
Urine
Declaration of Conformtiy RapidTrace
Tags: Certificates
User report: RapidTrace. The Laboratory of Racing Chemistry (LRC) analyzes samples from racehorses for doping drug. In order to screen approximately 1000 samples per week, LRC use an array of Biotage RapidTrace automated SPE systems. In an interview, Mr. Kenji Kinoshita and Mr. Kensaku Shirai talked about their experiences with the RapidTrace system in processing the heavy workload.
Tags: Customer Cases
A GC-MSMS method of simultaneously measuring legacy/emerging POPs
PCBs, OCPs and PBDEs) in serum was developed and applied to this study of exposure of mothers to harmful environmental contaminants.
Tags: BHC,
Clinical,
DDE,
DDT,
English,
Environmental,
GC-MS,
HCB,
Organochlorine pesticides (OCP),
Organochlorines,
Oxychlordane,
PBDE,
PBDEs,
PCB,
PCBs,
Persistant organic pollutants (POP),
Posters,
RapidTrace,
Serum,
Trans-nonachlor
This poster describes an automated procedure for isolation and clean up of 15 F2t-isoprostane (8-iso Prostaglandin 2a)(a lipid marker of oxidative stress) from BSA.
Tags: 15 F2t-isoprostane,
8-iso Prostaglandin2a,
BSA,
Biological fluids,
Biomarker,
Clinical,
Column,
English,
GC-MS,
Lipids,
Posters,
RapidTrace
In order to keep your RapidTrace® workstation clean, in good working order and free from contamination it is of course essential to
routinely clean and purge the fluid path. Whichever matrix and reagents are running through the system, traces of these liquids will
be left after use; proteins from plasma, precipitates from buffers or even particulates from oral fluid. Over time if the following
cleaning schedule is not observed these contaminants could cause blockages leading to malfunctions and/or cross contamination.
Tags: Product Notes
Every year, Biotage awards an employee that has come up with the best idea resulting in a successful new product or product improvement that makes it possible for Biotage to offer their custom...
13 January 2014Following the transplant of a heart, liver, kidney or lung, patients must be regularly tested for cyclosporine during the rest of their lives. Cyclosporine is an immunosuppressant us...
18 October 2011October 18, 2011 Uppsala, Sweden – Biotage® (STO: BIOT), a leading manufacturer of tools and technology for analytical and medicinal chemistry, introduced the new RapidTrace®+ SPE Wor...
Question:
My SPE technique is RapidTrace+. Is it possible to apply bigger sample volume (1000 mL) without prior evaporation?
Answer:
The maximum sample volume in standard operational mode is 40 mL limited to a 5.8 mL load each time. This is limited by the sample vial size and will only give you a 5 sample capacity.
Loading 40 mL of sample will be achieved by 6 x 5.8 mL then an additional 5.2 mL.
If you were to use one of the reagent lines as a source and use the condition function then it is possible to load more sample contained in an external vessel onto the SPE column via the reagent line. You will only be able to process a limited number of samples as dictated by other parameters within your SPE method as they will use reagent lines. Additionally you will not be able to load 1000 mL as you are limited by programming lines in a particular extraction procedure.
Essentially the RapidTrace is not really the right tool for the job. You would be better off considering the Biotage® Horizon 5000 depending on your application http://www.biotage.com/product-page/biotage-horizon-5000
Browse more questions and answers >This product was withdrawn on April 1st 2015. It has been replaced by the EVOLUTE® EXPRESS product line.
ISOLUTE® SLE+ Supported Liquid Extraction plates and columns are designed to provide stress free extraction of analytes from biological fluids, with high analyte recoveries, and clean, protein and phospholipid free extracts. Methods are easy to develop, and automation is simple.
Solid Phase Extraction (SPE) is a sample preparation technique used to isolate (purify) and/or concentrate (trace enrich) analytes. The range of possible analytes and matrices is vast. The ISOLUTE® range of SPE sorbents offers the scientist the breadth of choice needed to develop an optimised solution to a multitude of challenging sample preparation problems.
Extrahera™ is a powerful, user friendly automation system for Supported Liquid Extraction (SLE), Solid Phase Extraction (SPE), Phospholipid Depletion (PLD) and Protein Precipitation (PPT) based methods, in plate or column formats.
RapidTrace®+ delivers advanced processing and the rapid development of rugged sample preparation methods. It is extremely reliable and precise, incorporating easy-to-use features for quick method optimization. RapidTrace+ has been successfully implemented in a variety of laboratories all over the world.
Advanced bioanalytical techniques such as LC/MS/MS and GC/MS meet the increasing demand for greater speed. In many laboratories, solid phase extraction (SPE) has then become the throughput limiting step. RapidTrace®+ eliminates manual sample processing bottlenecks so that analysts can realize the full benefits of today’s powerful analytical instruments. RapidTrace+ is the perfect compliment to highspeed analytical techniques, a powerful high-throughput workstation dedicated to unattended automated sample preparation.
The enhanced RapidTrace®+ broadens usability and increases instrument performance. RapidTrace+ is compatible with industry standard 1, 3 and 6 mL syringe type columns and is compatible with all Biotage ISOLUTE® and EVOLUTE® products in these formats.
Increased bed height enables the use of a wider range of SPE cartridges including all ISOLUTE® and EVOLUTE® bed masses allowing for flexibility in methodology.
A new 5 position rack for 40 mL scintillation vials allows for larger sample volumes to be processed, often desired for food and environmental applications.
Easily leave extracted samples unattended. Samples are freshly prepared when you return.
10 RapidTrace®+ systems can be controlled by a single computer. The Modular design allows users to start with one module and expand up to 10 linked units. Modules within a group can run different methods independently and if one module is shut down for operator attention, other units remain operational so that sample throughput is not halted.
Utilizing the easy-to-use interface, it is simple to implement a structured, rational strategy for optimum SPE conditions in experiments.
Methods stored in the RapidTrace®+ controller can be instantly recalled and transferred to other units with 100% confidence.
Positive pressure displacement allows for accurate processing and improves reproducibility in flow rate sensitive methods.
Enclosed working parts ensures operator safety and reduced exposure to solvents and hazardous compounds. Additionally, password protected control software limits access to methods ensuring method security. Combined with magnetically encoded racks, run initiation is simplified and the possibility of loading an incorrect procedure is reduced.
Allows for more flexible use of solvents and provides an excellent system for method development.
Three segregated waste lines allows for separate aqueous, organic and biological waste disposal to simplify hazardous waste handling and reduce disposal costs.
Simple, rugged design with a minimum of moving parts and easily accessible consumables allows for a robust system with minimum “down-time”.
Utilizing the familiar Microsoft® Windows® interface, operators can easily implement a structured, rational strategy for method-development to determine optimum processing conditions for each sample. The methods can be written in minutes and recalled in seconds. Optimization of processing conditions is possible in a fraction of the time compared with conventional manual techniques. Column formats, reagents, concentrations and flow rates can be thoroughly explored as the system allows for incremental manipulation of any parameter (eg. pH, solvent concentration, cartridge type, flow rate etc)in order to determine effects on recovery and precision. Operators can insert, delete or manipulate steps and easily adjust parameters.
Methods can be assigned to an individual sample or by module. The methods stored in the RapidTrace+ controller can be easily transferred to other RapidTrace+ workstations. Additionally, methods can be saved and run concurrently or true, unattended, “multi-method” sample processing. This unique and rational method design capability dramatically speeds the development of rugged and reliable methods.
The RapidTrace+ Start-up Kit and Software, includes a Validation Manual that allows operators to walk through the validation process of their RapidTrace. This ensures that RapidTrace®+ provides the consistently clean extracts required for today’s sophisticated analytical instruments and regulatory bodies. The Validation Plan defines the methods and documentation utilized in the development, execution and reporting of a validation for a RapidTrace Workstation. It addresses Installation, Operation, and Performance Qualification. This comprehensive document enables operators to meet GLP, GMP and GOP requirements.
Magnetically Encoded Racks enable the user to designate a specific method to a rack. When using magnetically encoded racks, the method does not have to be initiated via the controller software, because the module reads the magnets and automatically knows which method to run. The module runs the number of samples corresponding to the number of sample test tubes placed in the rack. This means that software control of the system can be bypassed and methods will not be altered during routine operation. Labs around the world find this useful in production mode.